B.38 [SOURce:]CURRent:TRANsient COMMAND

CURR:TRAN

Syntax:

Short Form: [SOUR:]CURR:TRAN ,

Description: Configures a transient current pulse that is initiated by SYST:MODE:TRAN (see PAR. B.100).

See PAR. B.106 for details on using both Transient and Ramp commands at the same time. Figure B-
3 on page B-20 shows tra
nsient and ramp timing.

B.39 [SOURce:]CURRent:TRANsient? QUERY

CURR:TRAN?

Syntax:

Short Form: [SOUR:]CURR:TRAN?
Long Form: [SOURce:]CURRent:TRANsient?
Returns: ,

Description: Returns the transient current pulse parameters established by CURR:TRAN (see PAR. B.38).

B.40 [SOURce:]MODE COMMAND

MODE

Syntax:

Short Form: [SOUR:]MODE

Long Form: [SOURce:]MODE

where: = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.

Description: Sets operating mode: When the operating mode is changed, the input is set to off. Send INPut ON to

apply the new mode setting to the load.

B.41 [SOURce:]MODE? QUERY

MODE?

Syntax:

Short Form: [SOUR:]MODE?

Long Form: [SOURce:]MODE?

Returns: = CURR, POW, VOLT, RES, COND, SHORT or OFF.

Description: Indicates operating mode.

B.42 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] COMMAND

POW

Syntax:

Short Form: [SOUR:]POW[:LEV][:IMM][:AMP]
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]
Where = power setpoint

Description: Establishes power setpoint. The power setpoint only affects the load when the load is in Power

mode of operation.

B.43 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]? QUERY

POW?

Syntax:

Short Form: [SOUR:]POW[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]?
Returns = power setpoint

Description: Indicates power setpoint.

B.44 [SOURce:]POWer:PROTection COMMAND

POW:PROT

Syntax:

Short Form: [SOUR:]POW:PROT

Long Form: [SOURce:]POWer:PROTection

where = power protection setpoint

Description: Selects power protection setpoint. This is a user-defined value which, if exceeded during operation,

causes a FAULT condition, forcing the load to disengage.

B.45 [SOURce:]POWer:PROTection? QUERY

POW:PROT?

Syntax:

Short Form: [SOUR:]POW:PROT?

Long Form: [SOURce:]POWer:PROTection?

Returns = power protection setpoint

Description: Indicates power protection setpoint.