B.50 [source:]resistance:transient command, B.51 [source:]resistance:transient? query, B.46 – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.87 to 3.xx User Manual
Page 72: Source, B.47, B.48, B.49, B.50, B.51, B.52

B-10
SERIES EL 022013
B.46 [SOURce:]POWer:PROTection:STATe[:LEVel] COMMAND
POW:PROT:STAT 0
Syntax:
Short Form: [SOUR:]POW:PROT:STAT[:LEV] 0
Long Form: [SOURce:]POWer:PROTection:STATe[:LEVel] 0
Description: Clears power protection error.
B.47 [SOURce:]POWer:PROTection:STATe[:LEVel]? QUERY
POW:PROT:STAT?
Syntax:
Short Form: [SOUR:]POW:PROT:STAT[:LEV]?
Long Form: [SOURce:]POWer:PROTection:STATe[:LEVel]?
where:
0 (power protection error has not occurred)
Description: Indicates whether a power protection error has occurred (1) or not (0).
B.48 [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude] COMMAND
RES
Syntax:
Short Form: [SOUR:]RES[:LEV][:IMM][:AMP]
Long Form: [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude]
Where:
Description: Establishes resistance setpoint. The resistance setpoint only affects the load when the load is in
Resistance mode of operation. Resistance setpoint is stored in non-volatile memory; value is restored
upon power-up.
B.49 [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude]? QUERY
RES?
Syntax:
Short Form: [SOUR:]RES[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude]?
Returns:
Description: Indicates resistance setpoint.
B.50 [SOURce:]RESistance:TRANsient COMMAND
RES:TRAN
Syntax:
Short Form: [SOUR:]RES:TRAN
Description: Configures a transient resistance pulse that is initiated by SYST:MODE:TRAN (see PAR. B.100).
See PAR. B.106 for details on using both Transient and Ramp commands at the same time. Figure B-
3 on page B-20 shows transient and ramp timing.
B.51 [SOURce:]RESistance:TRANsient? QUERY
RES:TRAN?
Syntax:
Short Form: [SOUR:]RES:TRAN?
Long Form: [SOURce:]RESistance:TRANsient?
Returns:
Description:Returns the transient resistance pulse parameters established by RES:TRAN (see PAR. B.50).
B.52 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] COMMAND
VOLT
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]
where:
Description: Establishes voltage setpoint. The voltage setpoint only affects the load when the load is in Voltage
mode of operation. Voltage setpoint is stored in non-volatile memory; value is restored upon power-
up.