B.115 STATus:OPERation[:EVENt] QUERY

STAT:OPER?

Syntax:

Short Form: STAT:OPER[:EVEN]?

Long Form: STATus:OPERation[:EVENt]?

Return Value:

Description: Indicates changes in conditions monitored by Operational Event Register (see Table B-3).

Returns the value of the Operation Event register. The Operation Event register is a read-only register
which holds (latches) all events that occur. Reading the Operation Event register clears it. (See exam-
ple, Figure B-10.)

B.116 STATus:PRESet COMMAND

STAT:PRES

Syntax:

Short Form: STAT:PRES

Long Form: STATus:PRESet

Description: Disables reporting of all status events. This command sets the Operation Enabled Register (Table

B-3) to 8193 and the Questionable Register to 255. (See example, Figure B-10.)

B.117 STATus:QUEStionable[:EVENt]? QUERY

STAT:QUES?

Syntax:

Short Form: STAT:QUES[EVEN]?

Long Form: STATus:QUEStionable[EVENT]?

Return Value: actual register value

Description: Indicates the latched condition of the Questionable Event register. Returns the value of the

Questionable Event register (see Table B-4). The Questionable Event register is a read-only register
which holds (latches) all events. Only bits 13 and 12 are latched in the Status Questionable Event reg-
ister. Bits 0 and 1 of the Status Questionable Condition Register are not latched in the power supply.
Reading the Questionable Event register clears it. (See example, Figure B-10.)

B.118 STATus:QUEStionable:CONDition? QUERY

STAT:QUES:COND?

Syntax:

Short Form: STAT:QUES:COND?

Long Form: STATus:QUEStionable:CONDition?

Return Value: actual register value

Description: Returns the value of the Questionable Condition Register (see Table B-4). The Questionable

Condition Register contains unlatched real-time information about questionable conditions of the
power supply. Bit set to 1 = condition (active, true); bit reset to 0 = condition (inactive, false). Bits 1 or
0 may be both be set, indicating the power supply is settling after a voltage change. (See example,
Figure B-10.)

B.119 STATus:QUEStionable:ENABle COMMAND

STAT:QUES:ENAB

Syntax:

Short Form: STAT:QUES:ENAB Long Form: STATus:QUESionable:ENABle

Description: Programs Questionable Condition Enable Register (see Table B-4).The Questionable Condition

Enable Register determines which conditions are allowed to set the Questionable Condition Register;
it is a mask for enabling

specific bits in the Questionable Event register that can cause the question-

able summary bit (bit 3) of the Status Byte register to be set. The questionable summary bit is the log-
ical OR of all the enabled bits in the Questionable Event register

.

Bit set to 1 = function enabled

(active, true); bit reset to 0 = function disabled (inactive, false)

.

(See example, Figure B-10.)

TABLE B-4. QUESTIONABLE EVENT REGISTER, QUESTIONABLE CONDITION REGISTER

AND QUESTIONABLE CONDITION ENABLE REGISTER BITS

CONDITION

NU

SINK

CE

VE

NU

SE

NU

TE NU CM VM

BIT

15

14

13

12

11 - 7

6

5-4

3

2

1

0

VALUE

32,768

16,359 8192 4096 2048 - 128

64 32-16

8

4

2

1

CE

Current Error

CM Current Mode Error
NU

Not Used

SE

Slave Error

SINK Power Supply absorbing energy from load
TE

Thermal Error

VE

Voltage Error