B.67 [SOURce:]LIST:SAMPle:CURRent COMMAND

LIST:SAMP:CURR

Syntax:

Short Form: LIST:SAMP:CURR average,value

Long Form: LIST:SAMPle:CURRent average,value

where:

= number of measurements = integer: 2, 4, 8, 16, 32, 64, 128 or 256

= programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.

Description: Samples output current. is either programmed current in Amperes, or programmed voltage

in Volts. can only be 2, 4, 8, 16, 32, 64, 128 or 256. Entering a number other than those

permitted results in the average being set to the next lowest permitted number. The actual number of

points added to the list is . The samples are taken within the window established by

age>. The timing of the first sample is not exact. The sample duration must be established by the

LIST:SET:SAMP command, otherwise a -221 ”Settings conflict” error will result

This command must be preceded by LIST:CURR, LIST:VOLT or LIST:SET:SAMPLE command, other-

wise a -221 ”Settings conflict” error will result. The entries on the dwell list must match the values on

the current or voltage list, otherwise a -226 “Not Same Length” error will result. Only eight samples

may be taken in a list. If more than eight LIST:SAMP:CURR or LIST:SAMP:VOLT commands are

included in the list, a -223 “Too Much data“ error results. Also, note that each sample uses one of the

points in the waveform table; if the total number of points in the waveform exceeds 3933, -223 “Too

Much data“ error results. (See Figures B-6 and B-7.)

B.68 [SOURce:]LIST:SAMPle:VOLTage COMMAND

LIST:SAMP:VOLT

Syntax:

Short Form: LIST:SAMP:VOLT average,value
Long Form: LIST:SAMPle:VOLTage average,value
where:
average = number of measurements = integer: 2,4,8,16,32,64,128 or 256
value = programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.

Description: Samples output voltage. is either programmed current in Amperes, or programmed voltage

in Volts. can only be 2, 4, 8, 16, 32, 64, 128 or 256. Entering a number other than those

permitted results in the average being set to the next lowest permitted number. The actual number of

points added to the list is . The samples are taken within the window established by

age>. The timing of the first sample is not exact. The sample duration must be established by the

LIST:SET:SAMP command, otherwise a -221 ”Settings conflict” error will result

This command must be preceded by LIST:CURR, LIST:VOLT or LIST:SET:SAMPLE command, other-

wise a -221 ”Settings conflict” error will result. The entries on the dwell list must match the values on

the current or voltage list, otherwise a -226 “Not Same Length” error will result. Only eight samples

may be taken in a list. If more than eight LIST:SAMP:CURR or LIST:SAMP:VOLT commands are

included in the list, a -223 “Too Much data“ error results. Also, note that each sample uses one of the

points in the waveform table; if the total number of points in the waveform exceeds 3933, -223 “Too

Much data“ error results. (See Figures B-6 and B-7.)

B.69 [SOURce:]LIST:SAMPle? QUERY

LIST:SAMP?

Syntax:

Short Form: LIST:SAMP?
Long Form: LIST:SAMPle?
Returns:,....[V|C]

Description: Returns a comma-separated list of numbers representing points, followed by V or C. The point

listings are determined by the settings of LIST:SAMP:CURR and LIST:SAMP:VOLT commands. The
sample(s) occur after the points indicated. The type of list is indicated by V (Voltage) or C (Current).
(See Figures B-6 and B-7.)

B.70 [SOURce:]LIST:SET:SAMPle COMMAND

LIST:SET:SAMP

Syntax:

Short Form: LIST:SET:SAMP value
Long Form: LIST:SET:SAMPle value
where value = sample pulse duration, between 0.00025 and 0.034 (Seconds)

Description: Establishes sample pulse duration. This command must be used when measuring a sample, but

must only be issued once, otherwise a -100 “command error: will result. (See Figures B-6 and B-7.)