beautypg.com

B.45 [source:]function:mode command, B.46 [source:]function:mode? query, B.47 [source:]function:mode:trigger command – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.3.05 to 4.07 User Manual

Page 166: B.41, Sour, B.42, B.43, B.44, B.45, B.46

background image

B-12

BOP-1K 031912

B.41 [SOURce:]CURRent[:LEVel]:PROTect:LIMit:POSitive COMMAND

CURR:PROT:LIM:POS

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:LIM:POS
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:LIMit:POSitive
where is between minimum (box) value (Figure 1-3) and 1% above rated (nominal) output
current.

Description: Establishes the maximum value possible for the positive protection limit for current. This may be set

up to 1% greater than rated current. Requires MEM:UPD (PAR. B.12) to save for power up. See PAR.
3.5.2.2 and Figure 3-16 for special programming considerations.

B.42 [SOURce:]CURRent[:LEVel]:PROTect:LIMit:POSitive? QUERY

CURR:PROT:LIM:POS?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:LIM:POS?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:LIMit:POSitive?
Returns:
Description:Identifies the maximum value possible for the positive protection limit for current.

B.43 [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude] COMMAND

CURR:TRIG

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Programs current value to be transferred to output by *TRG (trigger) commands. This command

can be used to reset many power supplies to preselected parameters by issuing a single *TRG com-
mand. Actual output current will depend on load conditions. If the value exceeds the maximum for the
model being programmed, error message -222,”Data out of range” is posted in output queue. Related
Commands:
CURR, *TRG, TRIG

B.44 [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]? QUERY

CURR:TRIG?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns the current value established by CURR:TRIG command.

B.45 [SOURce:]FUNCtion:MODE COMMAND

FUNC:MODE

Syntax:

Short Form: FUNC:MODE {VOLT | CURR | EXT}
Long Form: [SOURce:]FUNCtion:MODE {VOLT | CURR | EXTernal}

Description: Establishes the operating mode of the power supply. VOLT = Constant Voltage mode (CV). CURR

= Constant Current mode (CC). EXT = External. FUNC:MODE VOLT commands power supply to Volt-
age mode, FUNC:MODE CURR commands power supply to Current mode. Commanded mode estab-
lishes parameters (voltage or current) monitored for error conditions. Actual mode depends upon load
conditions. When commanded to Voltage mode, if load conditions cause the power supply to try to
exceed the current limit, the unit will flag an error condition. When commanded to Current mode, if
load conditions cause the power supply to try to exceed the voltage limit, the unit will flag an error con-
dition.

B.46 [SOURce:]FUNCtion:MODE? QUERY

FUNC:MODE?

Syntax:

Short Form: FUNC:MODE?

Long Form: [SOURce:]FUNCtion:MODE?

Return value: 0 (voltage mode) or 1 (current mode)

Description: Returns the operating mode of the power supply. 0 = Constant Voltage mode (CV). 1 = Constant

Current mode (CC).

B.47 [SOURce:]FUNCtion:MODE:TRIGger COMMAND

FUNC:MODE:TRIG

Syntax:

Short Form: FUNC:MODE:TRIG {VOLT | CURR}
Long Form: [SOURce:]FUNCtion:MODE:TRIGger {VOLT | CURR}

Description: Establishes the operating mode of the power supply when a TRIGger command is sent. VOLT =

Constant Voltage mode (CV). CURR = Constant Current mode (CC). FUNC:MODE:TRIG VOLT com-
mands power supply to Voltage mode when a TRIGger command is sent, FUNC:MODE:TRIG CURR
commands power supply to Current mode when a TRIGger command is sent.