B.72 STATus:QUEStionable:CONDition? QUERY

STAT:QUES:COND?

Syntax:

Short Form: STAT:QUES:COND?

Long Form: STATus:QUEStionable:CONDition?

Return Value: actual register value

Description: Returns the value of the Questionable Condition Register (see Table B-5). The Questionable

Condition Register contains unlatched real-time information about questionable conditions of the
power supply. Bit set to 1 = condition enabled (active, true); bit reset to 0 = condition disabled (inac-
tive, false). Bits 1 or 0 may be both be set, indicating the power supply is settling after a voltage
change. (See example, Figure B-6.)

B.73 STATus:QUEStionable:ENABle COMMAND

STAT:QUES:ENAB

Syntax:

Short Form: STAT:QUES:ENAB Long Form: STATus:QUESionable:ENABle

Description: Programs Questionable Condition Enable Register (see Table B-5). The Questionable Condition

Enable Register determines which conditions are allowed to set the Questionable Condition Register;
it is a mask for enabling

specific bits in the Questionable Event register that can cause the question-

able summary bit (bit 3) of the Status Byte register to be set. The questionable summary bit is the log-
ical OR of all the enabled bits in the Questionable Event register

.

Bit set to 1 = function enabled

(active, true); bit reset to 0 = function disabled (inactive, false)

.

(See example, Figure B-6.)

B.74 STATus:QUEStionable:ENABle? QUERY

STAT:QUES:ENAB?

Syntax:

Short Form: STAT:QUES:ENAB? Long Form: STATus:QUESionable:ENABle?
Return Value: actual register value

Description: Reads Questionable Condition Enable Register (see Table B-5). Power supply returns value of

Questionable Condition Enable Register, indicating which conditions are being monitored. Bit set to 1
= function enabled (active, true); bit reset to 0 = function disabled (inactive, false)

.

Related Com-

mands: STAT:QUES?. (See example, Figure B-6.)

B.75 SYSTem:BEEP COMMAND

SYST:BEEP

Syntax:

Short Form: SYST:BEEP

Long Form: SYSTem:BEEP

Description: Causes the unit to emit a brief audible tone.

B.76

SYSTem:COMMunication:SERial:ECHO

COMMAND

SYST:COMM:SER:ECHO

Syntax:

Short Form: SYST:COMM:SER:ECHO {ON | OFF}
Long Form: SYSTem:COMMunication:SERial:ECHO {ON | OFF}

Description: Enables (ON) or disables (OFF) echo mode (see PAR. 4.5.2.1) Sending ON causes all subsequent

characters to be echoed back. Sending OFF turns off the character echo after the next line terminator
character. The *RST command has no effect on echo status. See PAR. 4.1.2 and Figure 4-1 for spe-
cial programming considerations.

B.77

SYSTem:COMMunication:SERial:ECHO?

QUERY

SYST:COMM:SER:ECHO?

Syntax:

Short Form: SYST:COMM:SER:ECHO?
Long Form: SYSTem:COMMunication:SERial:ECHO?
Return Value: {ON | OFF}

Description: Identifies whether echo mode is active (ON) or disabled (OFF) (see PAR. 4.5.2.1).

B.78

SYSTem:COMMunication:SERial:PACE

COMMAND

SYST:COMM:SER:PACE

Syntax:

Short Form: SYST:COMM:SER:PACE {NONE | XON}
Long Form: SYSTem:COMMunication:SERial:PACE {NONE | XON}

Description: Enables (XON) or disables (NONE) data flow control via the serial interface (see PAR. 4.5.2.2).

See PAR. 4.1.2 and Figure 4-1 for special programming considerations.

B.79

SYSTem:COMMunication:SERial:PACE?

QUERY

SYST:COMM:SER:PACE?

Syntax:

Short Form: SYST:COMM:SER:PACE {NONE | XON}
Long Form: SYSTem:COMMunication:SERial:PACE {NONE | XON}
Return Value: {XON | NONE}

Description: Identifies whether data flow control via the serial interface is enabled (XON) or disabled

(NONE) (see PAR. 4.5.2.2). See PAR. 4.1.2 and Figure 4-1 for special programming considerations.