B.45 [SOURce:]LIST:VOLTage COMMAND

LIST:VOLT

Short Form: LIST:VOLT[:LEV] , , . . . (to max of 1002 data points)

Long Form: LIST:VOLTage[:LEVel] , , . . . (to max of 1002 data points)
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Adds the voltage value (in Volts) to list. This command sequentially adds LIST:VOLTage values to

the main channel List Data Table locations illustrated in Table B-2. LIST:CLE sets starting location to 0.
Next location indicated by LIST:VOLT:POIN? These locations correspond to the default sequence
(LIST:GEN DESQ) The maximum number of entries is 1002.

Since the input buffer of the BIT 4886

has a limit of 253 characters, multiple commands are necessary to complete the full 1002 entries of
the list. If LIST:CURR has any entries, an error message: -221,”Settings conflict” is posted in the error
queue. Related Commands: LIST:VOLT:POIN?, FUNC:MODE, LIST:CLE, *RST. (See example, Fig-
ures B-2 and B-3.)

B.46 [SOURce:]LIST:VOLTage? QUERY

LIST:VOLT?

Syntax:

Short Form: LIST:VOLT?

Long Form: LIST:VOLTage?

Return Value: , , . . . to

Description: Identifies the parameters (main channel) entered for the list. Starting at location established by

LIST:QUERy, returns comma-separated list of up to 16 values indicating the main channel parameters
entered. i.e., the contents of the main channel locations of Table B-2. Related Commands: LIST:
VOLT, LIST:QUERy. If LIST:CURR has any entries, an error message: -221,”Settings conflict” is
posted in the error queue. (See example, Figure B-3.)

B.47 [SOURce:]LIST:VOLTage:POINts? QUERY

LIST:VOLT:POIN?

Syntax:

Short Form: LIST:VOLT:POIN? Long Form: LIST:VOLTage:POINts?
Return Value: (0 to 1001)

Description: Identifies the total number of points in a list and the next location to be filled by LIST:VOLT

command. The LIST:VOLT pointer is initially at 0 via *RST or LIST:CLE. For each data point entered
by a LIST:VOLT command the list pointer is incremented If LIST:VOLT:POIN? returns 5, the
LIST:VOLT pointer is at 5 indicating there are 5 data points comprising the list (locations 0 though 4)
and location 5 is the next to be filled. If LIST:CURR has any entries, an error message: -221,”Settings
conflict” is posted in the error queue. Related Commands: LIST:VOLT. (See example, Figure B-3.)

B.48 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND

CURR

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Sets programmed current level at power supply output. This command programs output current to

a specific value; actual output current will depend on load conditions. If CURR:RANG is set to 4, any
data that exceeds 1/4 of I

MAX

causes error message -222 “data out range” to be posted to the error

queue (See example, Figure B-4.)

B.49 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] QUERY

CURR?

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]? MIN, MAX
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? MIN, MAX
Return Value: = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns either the programmed value, maximum value, or minimum value of current. The

CURR? query returns the programmed value of current. Actual output current will depend on load con-
ditions. The CURR?MAX query returns the maximum current allowed for a particular model. CURR?
MIN returns minimum current allowed for power supply (always negative). Related Commands:
CURR. (See example, Figure B-4.)