B.66 status:operation:condition query, B.67 status:operation:enable command, B.68 status:operation:enable? query – KEPCO BIT 4886 Operator Manual User Manual
Page 86: B.69 status:operation[:event]? query, B.70 status:preset command, B.65, Source, B.66, Statu, B.67
![background image](/manuals/367872/86/background.png)
B-18
BIT 4886 120413
B.65 [SOURce:]VOLTage:[:LEVel]:TRIGgered[:AMPlitude]? QUERY
VOLT:TRIG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value:
Description: Returns value established by VOLT:TRIG command representing voltage value to be pro-
grammed by *TRG or TRIG command. (See example, Figure B-5.)
B.66 STATus:OPERation:CONDition QUERY
STAT:OPER:COND?
Syntax:
Short Form: STAT:OPER:COND?
Long Form: STATus:OPERation:CONDition?
Return Value:
Description: Returns the value of the Operation Condition Register (see Table B-4). The Operation Condition
Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-6.)
B.67 STATus:OPERation:ENABle COMMAND
STAT:OPER:ENAB
Syntax:
Short Form:
STAT:OPER:ENAB
Long Form:
STATus:OPERation:ENABle
Description: Sets Operation Enable Register. The Operation Enable Register is a mask for enabling specific bits
in the Operation Event Register which will cause the operation summary bit (bit 7) of the Status Byte
register to be set. Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inac-
tive, false). The operation summary bit is the logical OR of all the enabled bits in the Operation Event
register. (See example, Figure B-6.)
B.68 STATus:OPERation:ENABle? QUERY
STAT:OPER:ENAB?
Syntax:
Short Form: STAT:OPER:ENAB?
Long Form: STATus:OPERation:ENABle?
Return Value:
Description: Reads Operation Enable Register (see Table B-4). Returns value of Operation Enable Register bits.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-6.)
B.69 STATus:OPERation[:EVENt]? QUERY
STAT:OPER?
Syntax:
Short Form: STAT:OPER[:EVEN]?
Long Form: STATus:OPERation[:EVENt]?
Return Value:
Description: Indicates changes in conditions monitored by Operational Event Register (see Table B-4).
Returns the value of the Operation Event register. The Operation Event register is a read-only register
which holds (latches) all events that occur. Reading the Operation Event register clears it. (See exam-
ple, Figure B-6.)
B.70 STATus:PRESet COMMAND
STAT:PRES
Syntax:
Short Form: STAT:PRES
Long Form: STATus:PRESet
Description: Disables reporting of all status events. This command sets all bits of the Operation Condition
(Table B-4) and Questionable Condition Registers to 0, preventing all status events from being
reported. (See example, Figure B-6.)
TABLE B-4. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,
AND OPERATION EVENT REGISTER BITS
CONDITION
NU
CC
NU
CV
NU
NU
NU
NU
BIT
15-11
10
9
8
7 - 6
5
4 - 1
0
VALUE
32,768 -
2048
1024
512
256
128 - 64
32
16 -2
1
CC -
POWER SUPPLY IN CURRENT MODE
CV -
POWER SUPPLY IN VOLTAGE MODE
NU