B.51 [source:]current:mode? query, B.52 [source:]current:[:level]range command, B.53 [source:]current:[:level]range? query – KEPCO BIT 4886 Operator Manual User Manual

Page 83: B.54 [source:]current:[:level]range:auto command, B.51, Sourc, B.52, B.53, B.54, B.55

background image

BIT 4886 120413

B-15

B.51 [SOURce:]CURRent:MODE?

QUERY

CURR:MODE?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:MODE?
Long Form: [SOURce:]CURRent[:LEVel]:MODE?
Return value: FIXED or LIST or TRANS

Description: Identifies active current mode. Returns LIST while list is being executed. Returns TRANSIENT after

CURR:MODE:TRAN command has been issued, but before CURR: or *TRG command executes the
transient. Returns FIXED while in fixed (default) mode of operation. Related Commands: LIST com-
mands. (Similar to VOLT:MODE?, see Figure B-3.)

B.52 [SOURce:]CURRent:[:LEVel]RANGe COMMAND

CURR:RANG

Syntax:

Short Form: [SOUR:]CURR[:LEV]:RANG
Long Form: [SOURce:]CURRent[:LEVel]:RANGe
= 1 (full scale) or 4 (1/4 scale)

Description: Allows the user to specify the operating range for control of output current. Range = 1 allows

control of output current from 0 to full scale. Range = 4 allows the full range of the 16 bit D to A con-
verter to control 1/4 of the full scale output current. This command will generate a command warning
error if the unit is not in current mode. It will be remembered until a func:mode command is processed.
(See example, Figure B-4.)

B.53 [SOURce:]CURRent:[:LEVel]RANGe? QUERY

CURR:RANG?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:RANG?
Long Form: [SOURce:]CURRent[:LEVel]:RANGe?

Description: Identifies programmed current range. Returns 1 (full scale) or 4 (1/4 scale) (See example, Figure

B-4.)

CURR:RANG:AUTO

B.54 [SOURce:]CURRent:[:LEVel]RANGe:AUTO COMMAND

Syntax:

Short Form: [SOUR:]CURR[:LEV]:RANG:AUTO
Long Form: [SOURce:]CURRent[:LEVel]:RANGe:AUTO
= 1 (on) or 0 (off)

Description: Allows the user to specify automatic range for control of output voltage or current. Upon power

up or upon receiving *RST, automatic ranging is selected. Automatic ranging can be turned off by
VOLT:RANG:AUTO 0, CURR:RANG:AUTO 0, VOLT:RANG 1 or 4, or CURR:RANG 1 or 4. When
automatic ranging is set, the correct range is automatically selected based on the programmed param-
eter. If the programmed parameter is more than 1/4 of nominal full scale, full scale is selected; if equal
to or less than 1/4 of full scale, the 1/4 scale range is selected. For example, if auto ranging is set for a
unit with 100V nominal output operating in voltage mode, a programmed voltage of up to 25.0V auto-
matically selects range to 1/4 scale, while programming anything above 25.0V selects full scale. Func-
tion is identical to VOLT:RANG:AUTO.

B.55 [SOURce:]CURRent[:LEVel]TRIGgered[:AMPlitude] COMMAND

CURR:TRIG

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Programs current value to be transferred to output by *TRG or TRIG commands. This command

can be used to reset many power supplies to preselected parameters by issuing a single *TRG or
TRIG command. Actual output current will depend on load conditions. If the value exceeds the maxi-
mum for the model being programmed, error message -222,”Data out of range” is posted in output
queue. Related Commands: CURR, *TRG, TRIG (See example, Figure B-5.)

B.56 [SOURce:]CURRent[:LEVel]TRIGgered[:AMPlitude]? QUERY

CURR:TRIG?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns the current value established by CURR:TRIG command. (See example, Figure B-5.)