3.3.1

MANUAL CALIBRATION

1. Initiate calibration by sending the SCPI command

SYSTem:PASSword:CENable DEFAULT

and then send

CAL:STATe 1

. The password (DEFAULT) has been set at the factory. If the

password has been changed from DEFAULT, substitute the correct password for the unit in
the

SYST:PASS:CEN

command. If the password has been forgotten, consult factory.

If voltage calibration is not needed, proceed to step 16 for current calibration.

2. Set the BOP to zero volts output under the 1/4 range by sending

CAL:LVOLT ZERO

. Con-

nect a Digital Voltmeter (DVM) to the BOP output to measure the power supply output. Send
the Command

CAL:DPOT 32

to increase the output voltage until the DVM no longer

changes in response to the command. Then send the Command

CAL:DPOT -32

to the unit.

Adjust A1R81 on BOP until the DVM reads the voltage specified in Table 3-1 for LOW
RANGE ZERO.

3. Set the BOP output voltage to zero by sending

CAL:VOLT ZERO

. Send the command

CAL:DPOT 1

to increase the voltage or

CAL:DPOT -1

to decrease the voltage until the

DVM reads the voltage specified in Table 3-1 for VOLT ZERO.

4. Set the BOP to maximum positive output voltage by sending

CAL:VOLT MAX

. Measure the

voltage output using a DVM of at least 7 digits in accuracy.

5. Decrease the voltage output by sending

CAL:DPOT -1

commands until the voltmeter reads

less than the rated output voltage specified in Table 3-1 for MAX OUTPUT. Then send

CAL:DPOT 1

and verify the voltage is slightly above the rated output voltage specified in

Table 3-1 for MAX OUTPUT

6. Referring to Table 3-1 (HIGH RANGE ACCURACY) for value and ± tolerance, send the com-

mand

CAL:DATA -10

to decrease the voltage until the voltage is less than the rated output.

Send the command

CAL:DATA 1

to increase the voltage or the command

CAL:DATA -1

to

decrease the voltage until the DVM reads the value specified in Table 3-1 for HIGH RANGE
ACCURACY MAX.

7. Set the BOP to maximum negative output by sending

CAL:VOLT MIN

. Continue to measure

the output of the supply using the DVM.

8. If the output is less negative than the value specified in Table 3-1 for HIGH RANGE ACCU-

RACY MIN, send

CAL:DATA -100

and verify the output is now more negative. If the value

is still not more negative, send

CAL:DPOT -1

to change the course adjustment and provide

a more negative output.

Send the command

CAL:DATA 10

to increase the voltage. Continue to send

CAL:DATA 10

until the voltage is less negative than the value specified in Table 3-1 for HIGH RANGE
ACCURACY MIN. Send

CAL:DATA -1

until the reading is within the limits specified in Table

3-1 for HIGH RANGE ACCURACY MIN.

If the

CAL:DPOT

command was used to adjust the negative output, send

CAL:VOLT MAX

to

the unit and repeat step 6 before proceeding to step 9.

9. Set the BOP to 1/4 scale (low) range positive output voltage by sending

CAL:LVOLT MAX

.

Output voltage should be as specified in Table 3-1 for LOW RANGE NOMINAL. Measure the
voltage output using a Digital Voltmeter of at least 7 digits in accuracy.