2 calibration procedure, Calibration procedure -11, Step 4 – KEPCO BIT 4886 Operator Manual User Manual

Page 33: Step 5, Steps 6, 7, Steps 8, 9, Ep 3

background image

BIT 4886 120413

3-11

3.3.2.2

CALIBRATION PROCEDURE

1. Click on the Calibration Off button on the main panel (see Figure 3-4).This will open a cali-

bration panel (see Figure 3-4 and Table 3-4) that allows calibration in either voltage or cur-
rent mode.

2. The password (DEFAULT) has been set at the factory. If it has not been changed, it will be

displayed in the box adjacent to the Start button on the right side of the panel. If the pass-
word has been changed from DEFAULT, substitute the proper password in the box adjacent
to the Start button and click the Start button. If the password has been forgotten, consult
factory. If the password is correct, the Volt and Curr buttons will become active. Otherwise
the message

ERROR! - Wrong Password

is displayed and you must enter the correct

password and click Start again.

3. For Voltage Calibration, click Volt button. The text window reads

Volt ZERO

and the mes-

sage Adjust for ZERO R81 (Eo Zero) on BOP A1 card. is displayed.

4. Connect a DVM to the output of the unit and adjust the R81 Potentiometer on A1 board of

BOP until the DVM reads the voltage specified in Table 3-1 for Low Range Zero, then click
OK button.

NOTE: Use external DVM for all measurements.

5. The text window reads

Volt ZERO

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust the output until the

DVM reads the voltage specified in Table 3-1 for VOLT ZERO, then click OK button.

6. The text window reads

Volt MAX

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM reading is

as specified in Table 3-1 for High Range Accuracy. Click OK when done.

7. The text window reads

Volt

MIN

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM reading is

as specified in Table 3-1 for High Range Accuracy, then click OK button. NOTE: If the <<<—
or —>>> button is depressed, return to step 6 after clicking OK.

8. The text window reads

LVolt MAX

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust output until DVM

reads voltage specified in Table 3-1 for Low Range Accuracy, then click OK button.

9. The text window reads

LVolt MIN

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM reads volt-

age specified in Table 3-1 for Low Range Accuracy, then click OK button. NOTE: If the <<<—
or —>>> button is depressed, return to step 8 after clicking OK.

10.The text window reads

Volt PROTECT

and the message Measure and Adjust. is dis-

played. Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM

reading is as specified in Table 3-1. Click OK when done.

11. Voltage calibration is complete. If desired, the voltage calibration can be repeated by clicking

Volt at the calibration panel (Figure 3-4). If current calibration is not necessary, proceed to
step 23.