Sentinel CXT

Sentinel CXT Instruction Manual

Rev. 1.5

Page 24 of 42

4. Swipe the ENTER marker to initiate the search. The CXT controller will perform an incremental

search for available 700/100 and 600 serial sensors starting with Modbus™ address 01h and analog
inputs with a minimum current of 1.80mA.

NOTE

All sensors must be powered on to be detected correctly.

NOTE

The channel tags for the analog inputs will be set to “PPM H2S” by default. These can be
changed in the Utilities menu (Section 10.2).

NOTE

The channel tags for detected serial sensors will be set to gas type and units by default.
These can be changed through the Utilities menu.

5. The search will continue until Modbus™ address 7Fh is reached or until it is terminated by swiping

the ENTER marker which will save the serial sensors. When all of the connected sensors are
identified, swipe ENTER to begin detecting alarm stations with Modbus™ address 80h.

6. The search will continue until Modbus™ address 8Fh is reached or until it is terminated by swiping

the ENTER marker which will save the alarm stations. When all of the connected alarm stations are
identified, swipe ENTER to begin detecting wireless radios with Modbus™ address 01h.

7. The search will continue until Modbus™ address 7Fh is reached or the search is terminated by

swiping the ENTER marker which will save the wireless radio modules found.

NOTE

A swipe of the PROG marker at any time while in search mode will cancel the search and
restore the default settings. Any devices found up to that point will not be saved.

5.4.2

Set Channel Alarms

The Set Channel Alarms menu allows configuration of each channel’s alarms. The information to be
configured is the;

 ALM1 LEVEL,

 ALM1 ASCENDING,

 ALM2 LEVEL,

 ALM2 ASCENDING,

 ALM3 LEVEL, and

 ALM3 ASCENDING.

1. Upon entering this menu, the LCD will display:

CHANNEL 1 ALARMS:

ALM2 LEVEL:

##

ALM2 ASCENDING: X

ALM1 LEVEL:

##

ALM3 LEVEL:

##

ALM1 ASCENDING: X

ALM3 ASCENDING: X

2. A flashing cursor will appear on ALM1 LEVEL. Swiping the magnet over:

 The markers of the up or down arrows will move the flashing cursor to the desired function.

 The ENTER marker will select the function indicated by the flashing cursor

 The markers of the up or down arrows will change the value.

 The ENTER marker will save the selected value and return to the flashing cursor.