Emo-1005 troubleshooting – AW Gear Meters EMO-1005 User Manual
Page 39

AW-Lake Company 8809 Industrial Drive, Franksville, WI 53126 web: www.awgearmeters.com
Tel: 262-884-9800 Fax: 262-884-9810 Email: [email protected]
REV. 1a 05/98 EMO-1005 Manual.DOC
38
CAUTION!!!
The Power-supply card has live 110 VAC on the card.
Disconnect the power cord before opening the EMO-1005.
EMO-1005 TROUBLESHOOTING
Symptom
Possible Cause
Correction
No display at all.
No 110 VAC power to unit
Check 110 VAC Supply
Blown main fuse in power supply All
Check fuse F1
three LED
=s on supply card OFF
No display at all & Analog Output
+12 VDC/+5VDC supply fuse blown LED
Check fuse F2
is negative
D1 and D3 off
Analog 4-20 mA Output very High
-12 VDC supply fuse blown, LED D2 OFF
Check fuse F3
Analog 5 V Output very High
-12 VDC supply fuse blown, LED D2 OFF
Check fuse F3
Zero reading on rate or total. LED on
channel card is blinking.
KFR or KFT set to Zero
KFR and KFT must NOT be Zero
Correct KFR and KFT
Default KFR=100,
KFT=10000
Zero reading on display. LED on channel
card is NOT blinking.
Faulty pick-up, wiring wrong
Check pickup, Check wiring
Frequency amplitude too low
Wire from pickup is too long
No frequency input reading on pin 4.
LED on channel card is NOT blinking.
Frequency amplitude too low
Try pin 3 input.
Pin 4 amplitude must be at
least 4.5 V p-p
No frequency input reading on pin 3
LED on channel card is NOT blinking.
Sensitivity adjustment wrong.
Check jumper J10-J12 on
channel card.
Zero reading of both Rate and Total.
Sample amount set to Zero
KFR set to Zero
Set Sample amount to more
than Zero (Default=8)
Wrong reading of Rate or Totalizer.
KFR or KFT programmed wrong
Check formula for KFT or
KFT
Check programmed value
No Analog Input reading.
Faulty wiring or sensor KFA set to Zero
Check Status display, if NOT
OK, check sensor/wiring.
Check with mA meter on pin
5. If NO reading, check
wiring/sensor.
No Analog Output signal.
Faulty wiring/receiver
Check wiring/receiver. Check
with Volt meter between pin
6 and pin 7. Over 12 VDC =
Output OK, but open current
loop (1-5 VDC = 4-20 mA)
None of the Limit outputs are working
Common connection on pin 15 missing/bad
Check wiring
Lost configuration Data
Display
ΑRam and no Bat≅ at power-up
RAM battery dead
Return to AW-Company for
repair