STATUS DISPLAY MODE
Press the following key sequence:
CHA
->
DP
->
ENT
->
KFR
The status display will be displayed in the following form:
LINE 1
SDxx:XXXXPO/YYYY
LINE 2
AOzzzzAIyyyLOOIO
where,
xx: is the channel selected in line 1
XXXX is the input rate (frequency) in Hz
P indicates the mode: P = PID R = Complex Ratio
M = Monitor T = Transparent
G = Gun-On C = Closed Loop Control
O
indicates the Linearizer, O= Not active, L=Active
/ separator
YYYY is the totalizer lowest 4 digits in impulses
AO stands for Analog Output (1024 steps). 51.2 steps/mA.
zzzz is the current Analog Output value in decimal
(range is 0 - 1023)
AI stands for Analog Input (256 steps). 12.8 steps/mA.
yyy is the 8-bit analog converter value (000 - 255)
L stands for Limits, O = Limit On, I = Limit Off
OOIO
^.. Limit 1 Output (ON) (terminal 14)
^... Limit 2 Output (OFF) (terminal 13)
^.... Limit 3 Output (ON) (terminal 12)
^..... Limit 4 Output (ON) (terminal 11)
(Terminal 15 is common for Limits)
The numbers for Digital input or Analog input/output are unscaled numbers. Rate Scaling or gain
factors will have no effect on these numbers. By pressing the CHA key and then number 1, 2 or 3....
the status of other channels can be displayed. By doing this the status of the channel in line 1 of the
normal operating screen will be displayed on line 1. To go out of the Status display and back to the
normal operating display, press the ENT key. Check the status display first if there are any problems
with any inputs or outputs.
The Limits in the Status Display have no validity in PID mode.