beautypg.com

Using the menu bar – Yokogawa Value Series FX1000 User Manual

Page 35

background image

2-26

IM 04L21B01-63EN

Using the Menu Bar

On the menu bar, select [Window] - [Link] to display the [Link File] dialog box.

1. Click here

(display files for linking).

2. Select [Prev] (previous file), [Next] (subsequent file), or [All] (all files).

4. Displays the linked files.

3. Displays the linked files in a different color.

When the link settings file is saved (see the next page),

the extension .ldx is appended to the original file name.

Note

• When linking and displaying files, make sure that the number of data points after linking

does not exceed 5242880.

In addition, if there is a period over which data does not exist such as when a power failure

occurs, data is counted as if the data is acquired at the given scan interval even during that
period. The scan interval and the maximum period for linking files are indicated below.

Interval

Period

25 ms

36.4 hour

125 ms

7.5 days

1 s

60.6 days

10 s

606.8 days

For example, if data is captured continuously at a 125-ms interval and there is a one or

more week long power failure, the data from before and after the power failure cannot be
linked and displayed.

• If you open a file that was created prior to a time change on the recorder and then link

subsequent files, files from the open file up to the point of the time change are linked. If you
open a file that was created after a time change on the recorder and then link previous files,
files starting from the point of the time change are linked.

Starting (open) file

Time

Time

Recorder time changed

Starting (open) file

File created during the time change

File created during the time change

Files up to this file can be linked

Files starting from this file can be linked

• The linked data display is based on the time of the starting file.

2.5 Linking Files and Saving the Link Settings File