Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual
Page 75

5-18
<5. Actions of External Input and Output Signals >
IM 11B03A03-01E
(7)
Calibration factor
This register contains the calibration factor for each calibrated stream of the GC1000
Mark II. Since the factors are in the range of 0.000 to 9.999, each value is multiplied
by 1000 and displayed as integers, i.e., 0000 to 9999.
Table 5.5 Address Table (GC1000)
Description
Name
Address
Run command
Stop command
Time setting request
Calibration (validation)
command
Stream sequence select
command
Stream (Cont) select
command
001TT
0002P
00001
00002
Recieved the message by the master, the slave reset.
Same as above
Same as above
00003
0001P
Same as above P: Stream Sequence number (1 to 4)
Same as above P: 1 to 3 (Cal. 1 to 3), 4 to 6 (Val. 1 to 3)
TT: Stream number (01 to 31)
Analyzer normal
10001
Reset when the alarm status is read after alarm status change is read.
Analyzer error
10002
Analyzer status change
10003
Measuring
10004
Stop
10005
Maintenance
10006
Stream
sequence
1001P
Stream
(Cont)
10021
Calibration (Validation)
10022
Stream sequence
10023
Data update
101TT
Calibration factor update
102TT
Reset when the analysis value is read after data update is read. (TT:
Stream number)
Reset when the calibration factor is read after calibration factor
update is read. (TT: Stream number)
Data valid
11CCC
CCC: Component number
Alarm status
12AAA
AAA: Alarm number (001 to 249)
Year (40001), month/day (40002), hour (40003),
minute/second (40004)
Time setting request
40001 to 40004
Read only (Write disabled)
DDD(real format)5CCC*2-1(CCC: Component number)
Analysis value
41DDD
NN: Stream number +40, QQ: Peak number
Range select
4NNQQ
Stream number
30001
Starting peak number
301TT
Starting peak number of each stream (TT: Stream number)
Peak number of each stream (TT: Stream number)
Hour/Minute (TT: Stream number)
Seconds
Analysis value (CCC: component number)
Peak number
302TT
Sampling time
303TT
Retention time
32CCC
Calibration factor
33CCC
Analysis value
31CCC/-
31DDD
Analysis value (CCC: component number)
DDD(real format): CCC*2-1(CCC: Peak number)
P: Stream Sequence number (1 to 4)
T0505.EPS
3rd Edition : Aug. 23. 2006-00