Precautions Against Electrostatic Problems
The GC1000 system uses numerous IC components. When handling cards with IC
components mounted on them for maintenance or setting changes, take full precau-
tions against electrostatic problems.
These precautions are summarized below.
(a)
When storing or carrying cards, enclose them in a conductive bag or antistatic
bag. (Cards as shipped by Yokogawa are enclosed in a conductive bag or
antistatic bag labeled with cautions against electrostatic problems.)
(b)
Whenever mounting or demounting cards into or from a product, wear a wrist
strap grounded via a 1 M
Ω
resistance. Connect the wrist strap to any ground
terminal near the ground wire or to any unpainted part of the grounded frame.
Sheet
Gard
List strap
List strap
1M‰
F0003.EPS
Using a wrist strap and conductive sheet
Using a Conductive Sheet
(c)
When servicing cards on the bench, place them on a conductive sheet
grounded via a 1 M
Ω
resistance, wearing a wrist strap as in (2) above. Keep
easily-chargeable plastic materials away from the bench.
(d)
Never touch components mounted on the cards, the pattern side, connectors,
pin components, etc. with bare hands, unless using a wrist strap and a conduc-
tive sheet.
(e)
Wrist straps and conductive sheets are available from Yokogawa Engineering
Service (YSV).