P2 mean alarm, Source label – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 51

IBP2 Parameter Settings
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
45
P2 Mean Alarm
The following options allow you to set the limits of the P2 Mean alarm for all patient age
groups.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit parameter are:
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
Source Label
The Source Label parameter allows you to specify the label of the P2 channel.
The values for the Source Label parameter are:
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
Off
Possible Values:
-28-300
Default:
Adult: 120
Pediatric: 110
Neonate: 80
Possible Values:
-30-298
Default:
Adult: 50
Pediatric: 50
Neonate: 35
Possible Values:
P2, ABP, AO, BAP, CVP, FAP, ICP, LAP, PAP, RAP, UAP, UVP
Default:
P2