CPR Parameters
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
133
Enable Metronome Set to Always
When Enable Metronome is set to Always and Constant Metronome is Enabled, the metronome
is on when CPR is detected, and off when CPR is idle in all modes.
When Enable Metronome is set to Always and Constant Metronome is Disabled:
•
Manual Mode and Advisory/CPR Protocol Mode -- Metronome is only on compressions
when compressions are less than 80 per minute during CPR.
•
AED Mode -- Metronome is on at the start of CPR period.
Enable Metronome Set to AED Mode Only
When Enable Metronome is set to AED Mode Only and Constant Metronome is Enabled:
•
Manual Mode Advisory/CPR Protocol Mode- No metronome.
•
AED Mode -- Metronome is on when CPR is detected, and off when CPR is idle.
When Enable Metronome is set to AED Mode Only and Constant Metronome is Disabled:
•
Manual Mode Advisory/CPR Protocol Mode-- No metronome.
•
AED Mode -- Metronome is on at the start of CPR period.
The values for the Enable Metronome parameter are:
Table 1: CPR Metronome Activity
Possible Values:
Never, Always, AED Mode Only
Default:
Never
Enable
Metronome
Setting
AED Mode
Advisory/CPR
Protocol Mode
Manual Mode
Constant Metronome Enabled
AED Mode
Only
On when CPR detected
Off when CPR Idle
Off
Off
Always
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
Never
Off
Off
Off
Constant Metronome Disabled
AED Mode
Only
On at start of CPR period
Off
Off
Always
On at start of CPR period
On only at <80
compressions per minute
On only at <80
compressions per minute
Never
Off
Off
Off