ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 157

Supervisor Menu
9650-001358-01 Rev. B
ZOLL Configuration Manual
151
Alarms (cont’d)
Default Adult>RR/BR
Alarm (cont’d)
Lower Limit Enable
Off
On, Off
Upper Limit
50
5-150
Lower Limit
3
0-145
Default Adult>No
Breath Alarm
No Breath Alarm
Off
Off, 10 sec, 15 sec, 20 sec,
25 sec, 30 sec, 40 sec, 50
sec, 60 sec
Default Adult>SpO2
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
100
52-100
Lower Limit
85
50-98
Default Adult>SpCO
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
10
2-100
Lower Limit
0
0-98
Default Adult>SpMet
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
3.0
2.0-100.0
Lower Limit
0.0
0.0-98.0
Default Adult>T1
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
Default Adult>T2
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
Default Adult>
T
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Menu
Submenu
Parameter Name
Default Value
Parameter Values