Supervisor Menu
The following table includes parameter ranges and default values for the Supervisor menu:
Menu
Submenu
Parameter Name
Default Value
Parameter Values
Alarms
General
Default Patient Mode Adult
Adult, Pediatric, Neonate
Volume
Medium
Low, Medium, High
Alarm Suspend Time 4 min
Disable, 2 min, 4 min, 15
min, Always On
Default Adult>LTA
Monitoring
LTA Monitoring
Off
On, Off
Default Adult>HR/PR
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
120
32-300
Lower Limit
50
30-298
Default Adult>NIBP
Systolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
220
32-260
Lower Limit
75
30-258
Default Adult>NIBP
Diastolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
110
22-220
Lower Limit
35
20-218
Default Adult>NIBP
MAP Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
120
25-230
Lower Limit
50
23-228
Default Adult>EtCO2
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
60
5-150
Lower Limit
8
0-145
Default Adult>FiCO2
Alarm
Upper Limit Enable
Off
On, Off
Upper Limit
8
2-98
Default Adult>RR/BR
Alarm
Upper Limit Enable
Off
On, Off