ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 115

Defib/Pacer Parameters
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
109
Pediatric Defib Int. Pdl. Energy
The Pediatric Defib Int. Pdl. Energy parameter allows you to set the energy level of the internal
paddle for pediatric patients.
The values for the Pediatric Defib Int. Pdl. Energy parameter are:
Neonate Defib Int. Pdl. Energy
The Neonate Defib Int. Pdl. Energy parameter allows you to set the energy level of the internal
paddle for neonate patients.
The values for the Neonate Defib Int. Pdl. Energy parameter are:
Sync After Cardioversion
The Sync After Cardioversion parameter allows you to enable or disable synchronizing the
defibrillator discharge with the ECG R-wave.
The values for the Sync After Cardioversion parameter are:
Default Pacer Mode
The Default Pacer Mode parameter allows you to set the default Pacer Mode.
The values for the Default Pacer Mode parameter are:
Demand: pacing pulses are inhibited by the patients QRS complexes that occur during an
interval that is dependant on the setting of the rate control. If during that interval no QRS
complexes are detected, a pacing pulse is delivered to the patient, the number of which and the
rate are selected in the pacing rate window.
Fixed: pacing pulses are not dependant on the patient’s cardiac activity and are delivered at the
selected pacing rate.
Default Pacer Rate
The Default Pacer Rate parameter allows you to set the default Pacer rate.
The values for the Default Pacer Rate parameter are:
Possible Values:
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J, 10 J, 15 J, 20 J, 30 J, 50 J,
Default:
10 J
Possible Values:
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J, 10 J, 15 J, 20 J, 30 J, 50 J
Default:
10 J
Possible Values:
On, Off
Default:
Off
Possible Values:
Demand, Fixed
Default:
Demand
Possible Values:
30-180; <100 by 5’s, >100 by 10’s
Default:
70