Extreme brady/tachy alarms, Lta monitoring, Hr/pr tone – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 29: Hr/pr selected source

HR/PR Parameter Settings
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
23
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
Extreme Brady/Tachy Alarms
If LTA monitoring is enabled, the Extreme Brady/Tachy Alarms parameter allows you to set
the limits of the Extreme Brady/Tachy alarms.
The values for the Extreme Brady/Tachy Alarm parameter are:
LTA Monitoring
The LTA Monitoring parameter allows you to enable or disable LTA monitoring.
The values for the LTA Monitoring parameter are:
HR/PR Tone
The HR/PR Tone parameter allows you to enable or disable the HR/PR tone.
The values for the HR/PR Tone parameter are:
HR/PR Selected Source
The HR/PR Selected Source parameter allows you to specify the HR/PR source.
The values for the HR/PR Selected Source parameter are:
Possible Values:
30-298
Default:
Adult: 50
Pediatric: 50
Neonate: 100
Possible Values:
Tachy: +5-+30 by 5’s
Brady: -5--30 by 5’s
Default:
Tachy: +20
Brady: -10
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
On
Possible Values:
ECG, SpO2, P1, P2, P3
Default:
ECG