ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 166

160
zoll.com
9650-001358-01 Rev. B
Log (cont’d)
Treatment Options
(cont’d)
Treatment Button #8
MgSO4
O2, ASA, Nitro, Morph, IV, B
Block, Lido, MgSO4, Valium,
Sedate
Treatment Button #9
Valium
O2, ASA, Nitro, Morph, IV, B
Block, Lido, MgSO4, Valium,
Sedate
Treatment Button #10 Sedate
O2, ASA, Nitro, Morph, IV, B
Block, Lido, MgSO4, Valium,
Sedate
Waveform Recording Continuous
Waveform Recording
Disabled
Disabled, Enabled
Record Displayed
Waveforms
All Displayed
Waveforms
All Displayed Waveforms,
Top Waveform Only
Record Additional
Waveforms
CO2 Capnogram
CO2 Capnogram, IBP and
CO2 Capnogram
Record Pads Imp.
Waveform
Disabled
Enabled, Disabled
Defib/Pacer
Default Settings
Adult Defib Ext. Pdl.
Energy
120 J
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J,
8 J, 9 J, 10 J, 15 J, 20 J, 30
J, 50 J, 70 J, 85 J, 100 J,
120 J, 150 J, 200 J
Pediatric Defib Ext.
Pdl. Energy
50 J
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J,
8 J, 9 J, 10 J, 15 J, 20 J, 30
J, 50 J, 70 J, 85 J, 100 J,
120 J, 150 J, 200 J
Neonate Defib Ext.
Pdl. Energy
50 J
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J,
8 J, 9 J, 10 J, 15 J, 20 J, 30
J, 50 J, 70 J, 85 J, 100 J,
120 J, 150 J, 200 J
Adult Defib Int. Pdl.
Energy
10 J
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J,
8 J, 9 J, 10 J, 15 J, 20 J, 30
J, 50 J
Pediatric Defib Int.
Pdl. Energy
10 J
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J,
8 J, 9 J, 10 J, 15 J, 20 J, 30
J, 50 J
Neonate Defib Int.
Pdl. Energy
10 J
1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J,
8 J, 9 J, 10 J, 15 J, 20 J, 30
J, 50 J
Sync After
Cardioversion
Off
On, Off
Default Pacer Mode
Demand
Demand, Fixed
Default Pacer Rate
70
30-180; <100 by 5’s, >100 by
10’s
ECG
General
AC Mains Filter
60 Hz
Off, 50 Hz, 60 Hz
12 Lead ECG
Print on 12-Lead
Acquire
Off
On, Off
Print Copies
1
1, 2, 3, 4, 5
Menu
Submenu
Parameter Name
Default Value
Parameter Values