Parameter menus – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 150

144
zoll.com
9650-001358-01 Rev. B
Parameter Menus
The following table includes parameter ranges and default values for the Parameter menus:
Menu
Submenu
Parameter Name
Default Value
Parameter Values
HR/PR
HR/PR Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 120
Pediatric: 150
Neonate: 200
32-300
Lower Limit
Adult: 50
Pediatric: 50
Neonate: 100
30-300
Extreme Brady/Tachy
Alarms
Tachy: +20
Brady: -10
Tachy: +5-+30 by 5’s
Brady: -5--30 by 5’s
LTA Monitoring
Off
On, Off
HR/PR Tone
On
On, Off
HR/PR Selected
Source
ECG
ECG, SpO2, P1, P2, P3
ECG Sweep Speed
25 mm/s
25 mm/s, 50 mm/s
NIBP
NIBP Systolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 220
Pediatric: 145
Neonate: 100
Adult: 32-260
Pediatric: 32-160
Neonate: 22-120
Lower Limit
Adult: 75
Pediatric: 75
Neonate: 50
Adult: 30-258
Pediatric: 30-158
Neonate: 20-118
NIBP Diastolic Alarm Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 110
Pediatric: 100
Neonate:70
Adult: 22-220
Pediatric: 22-130
Neonate: 12-110
Lower Limit
Adult: 35
Pediatric: 35
Neonate: 30
Adult: 20-218
Pediatric: 20-128
Neonate: 10-108
NIBP MAP Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 120
Pediatric: 110
Neonate: 80
Adult: 25-230
Pediatric: 25-140
Neonate: 15-110