ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 161

Supervisor Menu
9650-001358-01 Rev. B
ZOLL Configuration Manual
155
Alarms (cont’d)
Default Pediatric>T2
Alarm (cont’d)
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
Default Pediatric>
T
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 5.0
°C: 2.8
°F: 0.2-90.0
°C: 0.2-50.0
Lower Limit
°F: 0.0
°C: 0.0
°F: 0.0-89.8
°C: 0.0-49.8
Default Pediatric>P1
Systolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
145
-28-300
Lower Limit
75
-30-298
Default Pediatric>P1
Diastolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
100
-28-300
Lower Limit
35
-30-298
Default Pediatric>P1
Mean Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
110
-28-300
Lower Limit
50
-30-298
Default Pediatric>P2
Systolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
145
-28-300
Lower Limit
75
-30-298
Default Pediatric>P2
Diastolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
100
-28-300
Lower Limit
35
-30-298
Default Pediatric>P2
Mean Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
110
-28-300
Lower Limit
50
-30-298
Menu
Submenu
Parameter Name
Default Value
Parameter Values