Temp Parameter Settings
The Temp parameters allow you to view and modify functions related to using the X Series unit
for temperature monitoring. These parameters include:
•
T1 Alarm
•
T2 Alarm
•
T
Alarm
•
T1 Source Label
•
T2 Source Label
T1 Alarm
The following options allow you to set the limits of the T1 alarm for all patient age groups.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit parameter are:
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
Off
Possible Values:
°F: 34.0-122.0
°C: 2.0-50.0
Default:
°F: 100.0
°C: 37.8
Possible Values:
°F: 32.0-120.0
°C: 0.0-48.0
Default:
°F: 95.0
°C: 35.0