Nibp mode, Nibp auto mode interval, Smartcuf on/off – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 34

Chapter 3
Parameter Menus
28
zoll.com
9650-001358-01 Rev. B
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
NIBP Mode
The NIBP Mode parameter allows you to set the NIBP Mode.
The values for the NIBP Mode parameter are:
Manual: Unit takes a single NIBP measurement when you press the NIBP button (
) on the
front panel. To repeat, press the NIBP key again.
Auto: Unit takes the first of a series of NIBP measurements when the NIBP Auto Mode
Interval timer expires, and then repeats the NIBP measurement at this specified interval.
NIBP Auto Mode Interval
If Auto Mode is selected, the NIBP Auto Mode Interval parameter allows you to set the interval
time.
The values for the NIBP Auto Mode Interval parameter are:
SmartCuf On/Off
The SmartCuf On/Off parameter allows you to enable or disable SmartCuf.
The values for the SmartCuf On/Off parameter are:
Possible Values:
Adult: 23-228
Pediatric: 23-138
Neonate: 13-108
Default:
Adult:
50
Pedi
atric: 35
Neonate: 35
Possible Values:
Manual, Auto
Default:
Manual
Possible Values:
1 min, 2 min, 3 min, 5 min, 10 min, 15 min, 30 min, 60 min
Default:
5 min
Possible Values:
On, Off
Default:
On