Trends parameters, Trend display format, Trend display interval – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 24: Trend on alarm

Chapter 2
Setup Menu
18
zoll.com
9650-001358-01 Rev. B
Trends Parameters
The Trends parameters allow you to view and modify settings related to trend information and
all monitored vital sign measurements. These parameters include:
•
Trend Display Format
•
Trend Display Interval
•
Trend on Alarm
Trend Display Format
The Trend Display Format parameter allows you to specify what vital sign measurements
appear on the Trends status monitor.
The values for the Trend Display Format parameter are:
Resp
Displays the following vital signs: HR, SpO2, RR, EtCO2, FiCO2
SpO2
Displays the following vital signs: HR, SpO2, SpCO
®
, SpMet
®
Temp
Displays the following vital signs: HR, SpO2, T1, T2,
T
P1
Displays the following vital signs: HR, SpO2, IBP1, RR
P2
Displays the following vital signs: HR, SpO2, IBP2, RR
P3
Displays the following vital signs: HR, SpO2, IBP3, RR
NIBP
Displays the following vital signs: HR, SpO2, NIBP, RR
Trend Display Interval
The Trend Display Interval parameter allows you to specify the time interval to display trends.
The values for the Trend Display Interval parameter are:
Trend on Alarm
The Trend on Alarm parameter allows you to specify whether or not to log Trends each time an
alarm occurs.
The values for the Trend on Alarm parameter are:
Possible Values:
Resp, SpO
2
, Temp, P1, P2, P3, NIBP
Default:
NIBP
Possible Values:
30 sec, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min
Default:
5 min
Possible Values:
On, Off
Default:
Off