ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 152

146
zoll.com
9650-001358-01 Rev. B
SpO2 (cont’d)
SpCO Alarm (cont’d) Lower Limit
Adult: 0
Pediatric: 0
Neonate: 0
0-98
SpMet Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 3.0
Pediatric: 3.0
Neonate: 3.0
2.0, 2.5, 3.0, 3.5, 4.0, 4.5,
5.0, 5.5, 6.0, 6.5, 7.0, 7.5,
8.0, 8.5, 9.0, 9.5, 10.0-100.0
Lower Limit
Adult: 0.0
Pediatric: 0.0
Neonate: 0.0
0.0, 0.5, 1.0, 1.5, 2.0, 2.5,
3.0, 3.5, 4.0, 4.5, 5.0, 5.5,
6.0, 6.5, 7.0, 7.5, 8.0, 8.5,
9.0, 9.5, 10.0-98.0
SpCO Monitoring
On
On, Off
SpMet Monitoring
On
On, Off
Averaging Time
8 seconds
4 seconds, 8 seconds, 16
seconds
Sensitivity
Normal
Normal, Maximum
HR/PR Tone
On
On, Off
Temp
T1 Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
T2 Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
T2 Alarm
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
T Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 5.0
°C: 2.8
°F: 0.2-90.0
°C: 02-50.0
Lower Limit
°F: 0.0
°C: 0.0
°F: 0.0-89.8
°C: 0.0-49.8
T1 Source Label
T1
T1, ART, CORE, ESOPH,
RECT, SKIN, VEN, NASO
T2 Source Label
T2
T2, ART, CORE, ESOPH,
RECT, SKIN, VEN, NASO
IBP1
P1 Systolic Alarm
Upper Limit Enable
Off
On, Off
Menu
Submenu
Parameter Name
Default Value
Parameter Values