Operation GC-Pro

TCD

Teledyne

Analytical

Instruments

80

4.5.12.9

C

YCLE

M

ODE

The detection cycle can be set for either a continuous measurement

cycle or for a single-shot measurement from the CYCLE MODE screen.
Although set at the factory for your application, the time allotment for
filling the sample chamber in the sample valve can also be set or
adjusted with this function.

From the SETTINGS menu, press CYCLE MODE. The following

screen appears:

Use the arrow in the Cycle Mode Selection field to toggle between

CONTINUOUS and MEASURE ONCE options. Then, if required, you
can alter the sample chamber charging time by entering appropriate
values in the FILL DURATION field.

Note: This has been set at the factory for your application and

should not require alteration. Contact TAI Customer
Service before making any changes to this parameter.

As with the other screens, you have the option of cancelling the entry
without saving changes using the CANCEL button and/or returning to
the previous screen with the BACK button.

4.5.13 Analysis Mode

Analysis mode is the default mode of the instrument where

measurements are updated after each cycle (forward + reverse) and
displayed on the HOME screen.