beautypg.com

Teledyne GC-Pro/TCD - Gas Chromatograph User Manual

Page 70

background image

Operation GC-Pro

TCD

Teledyne

Analytical

Instruments

56

From these screens (0-4 and 5-9) a specific group or groups can be

selected, names (tag) assigned, a range assigned as well as concentration
peaks (Sxx) added or subtracted.

A group is selected by touching the adjacent LED. When selected it

the LED icon should turn green.

A tag is limited to three characters (letters or numbers 0-9). A tag

can be assigned by touching any three keys provided in the on-screen
keypad after selecting the tag field for the relevant group. You must
press ENTER on the keypad to assign the tag before moving to another
setting or group, otherwise the tag will NOT be assigned.

To assign a range, touch the UP arrow button adjacent to the range

of the selected group.

Once all the required inputs have been made, select ENTER to save.

A confirmation window pops up to confirm the inputs or cancel.


Note: You can view the first of the Enabled Sxx on the Group

Setup screen for the respective Groups. If none of the Sxx
are enabled S** will be displayed. Details for enabling the
Sxx are given below.

It is possible to add or subtract concentration peaks to groups using

the SUM function available from the Group Select screen.

Selecting SUM from a selected group brings up an additional screen.