One the appropriate values have been entered select ENTER to save

the factor. A confirmation screen will arise asking the user to confirm
the changes before saving.

After confirmation, a display screen will appear where the user can

be the currently saved factor. If you “Cancelled” the entry on the last
screen, the previous Alt Span factor will be used.


To initiate an Alt Span calibration, touch “OK” to bring up the next

screen to select which group to span on or “CANCEL” to exit the Alt
Span function.

After pressing “OK” the next screen appears.

Use the “UP/DOWN” arrows to enter the group to span on and then

press “ENTER” to begin the calibration or “CANCEL” to exit the
function.

Once the calibration has completed the HOME screen will appear on

the display.