GC-Pro TCD

Operation

Teledyne

Analytical

Instruments

57

This screen has three inputs for ten peaks S00 through S09. The

possible inputs are Subtract, Add, and NA (Not Applicable). The user
can assign these inputs by touching the UP arrow button available. After
touching the ENTER button to save the inputs, a confirmation window
pops up.

Note: After selecting the ENTER button there will be a delay of

few seconds to update the inputs. All other buttons (i.e.
home, cancel, back etc) available on the screen will be
disabled during the delay period.

As an example, assume Groups G00 and G01 have been setup to

monitor two different hydrocarbons with peaks associated as follows:

Group

Assigned Peak

G00 S00

G01 S01