Teledyne GC-Pro/TCD - Gas Chromatograph User Manual

Page 39

background image

GC-Pro TCD

Installation

Teledyne

Analytical

Instruments

25

1. Test the CRC. When the CRC is open, send a span command

until the CRC closes (The CRC will quickly close.)

2. When the CRC closes, remove the span command.

When CRC opens again, the span calibration is done, and the sample

is being analyzed.

3.3.2.7

‘M

EASURE

O

NCE

(

OR ONE SHOT MEASUREMENT

)

CONTACTS

There are three dedicated relay contacts and these are available from

the bottom Equipment Interface Connector. They are assigned to
indicate status of ‘measure once’ (or one shot measurement). Contacts
are normally open, and they close when GC-Pro TCD switches to that
particular range.

Table 3-5: ‘Measure Once’ Relay Connections

Pin

Function

9

+ Start Measurement Digital Input

11

- Start Measurement Digital Input

21

IDLE Contact 1

38

IDLE Contact 2

22

IN MEASUREMENT Contact 1

39

IN MEASUREMENT Contact 2

19

MEASUREMENT DONE Contact 1

18

MEASUREMENT DONE Contact 2

The MEASUREMENT DONE contact only closes for 5 seconds at

end of the measurement cycle. After that, the IDLE contact closes. The
display reading and analog output will then remain fixed until a new
measurement is requested.

Digital input range is 5 to 24 vdc to start measurement.

3.3.2.9

P

IN

O

UT

T

ABLE

The following table summarizes all the outputs/inputs available on

the three 50 pin D-Sub connectors on the back panel of the analyzer.