Operation GC-Pro

TCD

Teledyne

Analytical

Instruments

54

You can adjust any of the three PID constants by touching the data

box and entering a new value into the numerical entry box that pops up.
Once the value has been entered, press ENTER on the entry box to
accept the value and dismiss the numerical entry box. After all
parameters have been set, save the values by pressing ENTER on the
lower left corner of the screen. This will return you to the Temperature
main screen.

4.5.7.4

S

TATUS

The STATUS screen displays the status for the three thermocouple

sensors. If a thermocouple is open or damaged, this condition will be
indicated by the message “Broken” in the corresponding sensor column.
Otherwise, if a thermocouple is operating properly “Normal” will be
displayed in the sensor column.