Operation GC-Pro

TCD

Teledyne

Analytical

Instruments

38

4.4 Activating the Support Gases

The instrument gas controls are located on the front panel adjacent to

the touch screen display as shown in Figure 1-1. Make sure that the
following support gases are available at the analyzer:

 Carrier gas: (usually nitrogen but refer to Appendix C for

your specific application.

 Span gas: Typically 70-90% of measured component on

the range of interest in a clean background gas.
Refer to Appendix C for the specific span gas
composition for your application.

4.4.1 Carrier Gas

Set the carrier gas source regulator to 80 psig and adjust the analyzer

sample regulator until the sample pressure gauge reads the
recommended sample pressure of 7.0 psig.

4.4.2 Span Gas

1. Feed span gas to the analyzer (see Section 4.4.2). Gas switching

can be performed manually or handled through the auto
calibration valves (if the option is added). See Section 4.5.10.3.

2. Observe that the analyzer sample flow meter reads from 0.3 to

1.0 SCFH.

3. Turn off span gas.

4.5 Analyzer Operation

Although the GC-Pro TCD has been programmed for your

application at the factory, it can be further configured at the operator
level. Depending on the specifics of the application, this might include
all or a set of the following procedures:

1. Setting system parameters

 Establish a security password, if desired, requiring operator

to log in.

 Establish and start an automatic calibration cycle (if equipped

with the optional auto calibration module).

2. Routine operation.