GC-Pro TCD

Installation

Teledyne

Analytical

Instruments

21

0–1 VDC Range ID:

0.20 V = Low Range
0.5 V = Medium Range
0.80 V = High Range

4–20 mA DC % Range: Current increases linearly with

concentration, from 4 mA at 0
concentration to 20 mA at full scale.
(Full scale = 100% of programmable
range.)

4–20 mA dc Range ID:

6.8 mA = Range 1
12.0 mA = Range 2
16.8 mA = Range 3

Table 3-1: Analog Output Connections — Middle Connector

Pin

Function

3

Channel 2, + 4-20 mA, floating

4

Channel 2, – 4-20 mA, floating

5

Channel 1, + 4-20 mA, floating

6

Channel 1, – 4-20 mA, floating

8

Channel 2, 0-1 VDC

23

Channel 2, 0-1 VDC, negative ground

24

Channel 1, 0-1 VDC

7

Channel 1, negative ground

Examples:

The analog output signal has a voltage which depends on gas

concentration relative to the full scale of the range. To relate the signal
output to the actual concentration, it is necessary to know what range the
instrument is currently on, especially when the analyzer is in the auto
ranging mode.

The signal output for concentration is linear over the currently

selected analysis range. For example, if the analyzer is set on a range
that was defined as 90-100% nitrogen then the output would be as
shown in Table 3-2.