beautypg.com

Teledyne GC-Pro/TCD - Gas Chromatograph User Manual

Page 34

background image

Installation GC-ProTCD

Teledyne

Analytical

Instruments

20

3.3.2.2

F

USE

I

NSTALLATION

The fuse block, at the right of the power cord receptacle, accepts US

or European size fuses. A jumper replaces the fuse in whichever fuse
receptacle is not used.

3.3.2.3

50-P

IN

E

QUIPMENT

I

NTERFACE

C

ONNECTOR

There are three 50-pin Equipment Interface Connectors installed on

the rear panel of the GC-Pro TCD. Figure 3-2 shows the pin layout of
the Equipment Interface connectors. The arrangement is shown as seen
when the viewer faces the rear panel of the analyzer. The pin numbers
for each input/output function are given where each function is
described in the paragraphs below.

Figure 3-2: Equipment Interface Connector Pin Arrangement

From top to bottom the signals handled by the individual connectors

are as follows:

 Top connector:

Alarm relay connections

 Middle connector: Analog output signals
 Bottom connector: Standard connector for all additional

signals

See Section 3.3.2.9 for Pin-Out connections for each connector.

3.3.2.4

A

NALOG

O

UTPUT

There are four DC output signal pins—two pins per output. These

are found on the middle connector. For polarity, see Table 3-1. The
outputs are:

0–1 VDC % of Range:

Voltage rises linearly with increasing
concentration, from 0 V at 0
concentration to 1 V at full scale. (Full
scale = 100% of programmable range.)