Teledyne GC-Pro/TCD - Gas Chromatograph User Manual

Page 37

background image

GC-Pro TCD

Installation

Teledyne

Analytical

Instruments

23

Threshold Alarm 2:

 Can be configured as high (actuates when

concentration is above threshold), or low (actuates
when concentration is below threshold).

 Can be configured as failsafe or non-failsafe
 Can be configured as non-latching
 Can be configured out (defeated).

System Alarm:

 Actuates when DC power supplied to circuits is

unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be
defeated.

Note: Reset by pressing the STANDBY button to remove power.

Then press STANDBY again and any other button except
SYSTEM to resume. Further detail can be found in
Chapter 4, Section 4.5.4.

Table 3-3: Alarm Relay Contact Pins — Top Connector

Pin Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact