beautypg.com

Teledyne GC-Pro/FID - Gas Chromatograph User Manual

Page 84

background image

Operation GC-Pro

FID

Teledyne

Analytical

Instruments

70

Pressing the “Settings” icon displays the following screen:

4.7.12.1

A

LARMS

The GC-Pro FID provides two concentration alarms with relays.

When an alarm condition exists, the relay or relays will change state.
The relays are available from the 50-pin Equipment Connector as
described in Section 3.3.2.5 and 3.3.2.9. The alarms screens are used to
configure how the concentration alarms operate.

There are two identical screens within the ALARMS function, one

for setting the parameters on ALARM 1 and the other for ALARM 2.

Pressing the ALARMS button brings up a screen where you can

change or view the alarm parameters for either Alarm 1 or Alarm 2.