beautypg.com

Appendix – Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual

Page 69

background image

Trace Oxygen Analyzer

Appendix

Teledyne

Analytical

Instruments

57

Appendix

A-1 Specifications

Packaging:

General Purpose

19” standard instrument rack mount.

Sensor:

Zirconium oxide sensor.

Output

Signal:

0-1 VDC % of range

0-1 VDC Range ID
4-20 mA DC % of range (optional)
4-20 mA DC Range ID (optional)

Sample System:

All wetted parts of 316 stainless steel.

90 % Response Time:

65 seconds at 25 °C (77 °F) on 10 ppm

range, 100 seconds on 1 ppm range.

Ranges:

Three programmable ranges from 0-1 ppm

to 0-100 % oxygen.

Auto ranging with range ID output.

Alarms:

One system-failure alarm contact to detect

power failure.

Two adjustable concentration threshold

alarm contacts with fully programmable
setpoints.

Displays:

2-line by 20-character, VFD screen, and

one 5 digit LED display.

Digital Interface:

Full duplex RS-232 communications port.

Power:

Factory configured for either 115 VAC or

230 VAC operation, 50/60 Hz.

Operating Temperature: 5-35 °C