Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual
Page 32
Installation 3000ZA2G
Teledyne
Analytical
Instruments
20
Table 3-1: Analog Output Connections
Pin
Function
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ ppm Range, 4-20 mA, floating
6
– ppm Range, 4-20 mA, floating
8
+ Range ID, 0-1 VDC
23
– Range ID, 0-1 V dc, negative ground
24
+ ppm Range, 0-1 VDC
7
– ppm Range, 0-1 VDC, negative ground
Alarm Relays:
The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The possible alarm configurations are as follows:
Threshold Alarm 1:
Can be configured as high (actuates when concentration is
above threshold), or low (actuates when concentration is
below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
Threshold Alarm 2:
Can be configured as high (actuates when concentration is
above threshold), or low (actuates when concentration is
below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).