Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual
Page 59
Trace Oxygen Analyzer
Operation
Teledyne
Analytical
Instruments
47
so that maintenance can be performed under conditions
which would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in
section 4.3.3 to enter your password. Once you have clearance to
proceed, enter the Alarm function.
Press the Alarm button on the front panel to enter the Alarm
function. Make sure that AL–1 is blinking.
AL—1 AL—2
Choose Alarm
Set up alarm 1 by moving the blinking over to AL–1 using the
◄►arrow keys. Then press Enter to move to the next screen.
AL—1 1000
ppm HI
Dft—N Fs—N Ltch—N
Five parameters can be changed on this screen:
Value of the alarm setpoint, AL–1 #### ppm (oxygen)
Out-of-range direction, HI or LO
Defeated? Dft–Y/N (Yes/No)
Failsafe? Fs–Y/N (Yes/No)
Latching? Ltch–Y/N (Yes/No).
To define the setpoint, use the ◄►arrow keys to move the
blinking over to AL–1 ####. Then use the ▲/▼ arrow keys to change the
number. Holding down the key speeds up the incrementing or
decrementing. (Remember, the setpoint units are ppm O
2
.)
To set the other parameters use the ◄►arrow keys to move the
blinking over to the desired parameter. Then use the ▲/▼ arrow keys to
change the parameter.