Operation 3000ZA2G


Teledyne

Analytical

Instruments

38

4.3.5 System Self-Diagnostic Test

The Model 3000ZA2G has a built-in self-diagnostic testing routine.

Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the
test the status of each function is displayed on the screen, either as OK
or as a number between 1 and 3. (See System Self Diagnostic Test in
Chapter 5 for number code).

The self diagnostics are run automatically by the analyzer

whenever the instrument is turned on, but the test can also be run by the
operator at will. To initiate a self diagnostic test during operation:

Press the System button to start the System function.

TRAK/HLD Auto—Cal

PSWD Logout More


Use the ◄►arrow keys to blink More, and then press Enter.

Version Self—Test

Use the ◄►arrow keys again to move the blinking to the Self–

Test function. The screen will follow the running of the diagnostic.
When the testing is complete, the results are displayed.

Power: OK

Analog: OK


The module is functioning properly if it is followed by OK. A

number indicates a problem in a specific area of the instrument. Refer to
Chapter 5 Maintenance and Troubleshooting for number-code
information. The results screen alternates for a time with:

Press Any Key

To Continue...