Introduction 3000ZA2G


Teledyne

Analytical

Instruments

6

Sample Inlet and Outlet: One inlet (must be externally
valved) and one exhaust. Three inlets when auto calibration
option is present.

Reference Inlet and Outlet: Flowing air reference inlet and
exhaust.

9-Pin RS-232 Port: Serial digital concentration signal
output and control input.

50-Pin Equipment Interface Port:

Analog Outputs: 0–1 VDC oxygen concentration plus
0-1 VDC range ID
Alarm Connections: 2 concentration alarms and 1
system alarm.
Remote Valve: Used in the 3000ZA2G for controlling
external solenoid valves only.
Remote Span/Zero: Digital inputs allow external
control of analyzer calibration.
Calibration Contact: To notify external equipment that
instrument is being calibrated and readings are not
monitoring sample.
Range ID Contacts: Four separate, dedicated, range
relay contacts. Low, Medium, High, Cal.
Network I/O: Serial digital communications for local
network access. For future expansion. Not currently
implemented.

Optional

Calibration Gas Ports: Separate fittings for zero, span, and
sample gas input, internal valves for automatic gas switching.

Current Signal Output: Additional isolated 4–20 mA DC
plus 4–20 mA DC range ID.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3000ZA2G is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 30-45 minutes
per day.