33


Press ◄►arrows to blink

Span

(or

Zero

), then press Enter again.

(You won’t be able to set OFF to ON if a zero interval is entered.) A
Span Every ... (or Zero Every ...) screen appears.

Span Every 0 d

Start 0 h from now


Use ▲/▼ arrows to set an interval value, and then use ◄ arrows

to move to the start-time value. Use ▲/▼ arrows to set a start-time
value.

To turn ON the Span and/or Zero cycles (to activate Auto-Cal):

Press System again, choose

Auto—Cal,

and press Enter again. When

the Span/ Zero values screen appears, use the ◄► arrows to blink the
Span (or Zero) OFF/ON field. Use

▲/▼ arrows to set the OFF/ON

field to ON. You can now turn these fields ON because there is a
nonzero span interval defined.

4.3.3 Password Protection

If a password is assigned, then setting the following system

parameters can be done only after the password is entered:

Span and zero settings

Alarm setpoints

Analysis range definitions

Switching between auto ranging and manual override

Setting up an auto-cal

Assigning a new password

However, the instrument can still be used for analysis or for

initiating a self- test without entering the password.

If you have decided not to employ password security, use the

default password TAI. This password will be displayed automatically by
the microprocessor. The operator just presses the Enter key to be
allowed total access to the instrument’s features.