beautypg.com

Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual

Page 29

background image

Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

17

3.3.1 Gas Connections

All gas and exhaust connections are made to the labeled 1/4” tube

fittings (6 mm adapters for metric system installations are available) on
the rear panel. The rear panel connections are different between the
standard unit and one which includes the auto calibration option. For
safe connections:

1. Insert the tube into the fitting and finger-tighten the nut until

the tubing cannot be rotated freely by hand in the fitting. This
may require an additional 1/8 turn beyond finger tight.

2. Hold the fitting body steady with a backup wrench. Using

another wrench, rotate the nut another 1-1/4 turns.

SAMPLE IN:

In the standard model, gas connections are made at the

SAMPLE IN

and

EXHAUST OUT

connections. Calibration gases must tee into the sample

inlet with appropriate valves (not supplied).

The inlet gas pressure should be reasonably regulated. Pressures

between 2 and 50 psig are acceptable as long as the pressure, once
established, will keep the front panel flowmeter reading in the
acceptable range.

Use the sample flow control valve on the front panel to adjust the

flow rate to within the acceptable range of 0.1-0.3 SLPM.

If greater sample flow is required for improved response time,

install a bypass in the sampling system upstream of the analyzer input.
SAMPLE VENT:
Exhaust connections must be consistent with the hazard level of the
constituent gases. Check local, state, and federal laws, and ensure that
the exhaust stream vents to an appropriately controlled area, if required.
REF IN/OUT:

Flowing air reference. Air must be clean, dry, and free of hydrocarbons.
Use the reference flow control valve on the front panel to adjust the flow
rate to within the acceptable range of 0.1-0.2 SLPM.

Note: It is possible to use the instrument without a flowing

reference however doing so will result in a marginal
reduction in performance and is not recommended for best
performance.