5.4 System Self Diagnostic Test

1. Press the SYSTEM button to enter the system mode.
2. Use the ◄► arrow keys to move to More, and press

ENTER.

3. Use the ◄► arrow keys to move to Self-Test, and press

ENTER.

The following Items are tested:

Table 5-1: Self-Test Result Display

Power: OK

5V and +/-15V power supplies

Analog: OK

Measurement and range ID DACs


If any item above fails this test, it will be listed on the VFD display

followed by a code number instead of “OK

5.5 Major Internal Components

The zirconium sensor assembly is accessed by removing the top

cover of the analyzer. The electronic PC boards are accessed by opening
the door on the front panel. The gas piping is illustrated in Figure 2-3
and the major electronic components locations are shown in Figure 2-4,
in Chapter 2.

CAUTION:

SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.


The 3000ZA2G contains the following major components:

Analysis Section

Zirconium oxide sensor assembly

Sample system

Power Supply

Microprocessor

Displays

5 digit LED meter