beautypg.com

Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual

Page 55

background image

Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

43

Calib. Holding time

Cal hold: 5 min


This menu allows the operator to set the time the analyzer should

be held in the span mode, after the readings of the analyzer settle. Five
minutes is the default, but it could be adjusted anywhere from 1 to 60
minutes by using the UP or DOWN keys.

Press Enter to move to the next screen.

Span Val: 000008.00

Span Mod #


Use the ▲/▼ arrow keys to enter the oxygen-concentration mode.

Use the ◄►arrow keys to blink the digit you are going to modify. Use
the ▲/▼ arrow keys again to change the value of the selected digit.
When you have finished typing in the concentration of the span gas you
are using, press Enter to begin the Span calibration.

#### ppm

Span

Slope=#### ppm/s


The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates
information on Slope. Spanning automatically ends when the span
output corresponds, within tolerance, to the value of the span gas
concentration. Then the instrument automatically returns to the analyze
mode.

4.4.2.2

M

ANUAL

M

ODE

S

PANNING

Press Span to start the Span function. The screen that appears

allows you to select whether the span calibration is to be performed
automatically or manually.